Line 1: Line 1:
 +====== Command:​AFLIM (AL) ======
  
 +--> For CRISP#
 +
 +--> Tiger Syntax#
 +
 +^Shortcut |AL|
 +^Format ​ |[Addr#]AL [X=Log_amp_AGC] [Y=LED_intesity_pot] [Z=in_focus_mm] ​  |
 +^Type |Card-Addressed|
 +^Remembered |Using [Addr#]SS Z |
 +
 +<--
 +
 +--> MS2000 and RM2000 Syntax#
 +
 +^Shortcut |AL|
 +^Format ​ |AL [X=Log_amp_AGC] [Y=LED_intesity_pot] [Z=in_focus_mm] ​  |
 +^Remembered |Using SS Z |
 +
 +<--
 +
 +X and Y arguments of this command to directly read and write values (0 to 255) to the CRISP electronics digital potentiometers. ​ (Not recommended for use with host software.)
 +
 +The Z-argument specifies the focus precision (in millimeters) when the lock state changes from K or k to F. Useful for automatic checking of desired focus stability. ​ Also useful to enforce a tighter or looser focus state before indicating a lock condition. ​ Note that this value is overwritten whenever the NA of the objective is specified via the LR Y command as of November 2015.
 +
 +<--
 +
 +--> For Video Autofocus#
 +
 +--> Tiger Syntax#
 +
 +^Shortcut |AL|
 +^Format ​ |[Addr#]AL [X= x-axis highlight] [Y= y-axis highlight] [Z= safety limit enable] ​ |
 +^Type |Card-Addressed|
 +^Remembered |Using [Addr#]SS Z |
 +
 +<--
 +
 +--> MS2000 and RM2000 Syntax#
 +
 +^Shortcut |AL|
 +^Format ​ |AL [X= x-axis highlight] [Y= y-axis highlight] [Z= safety limit enable] ​ |
 +^Remembered |Using SS Z |
 +
 +<--
 +
 +
 +
 +The X and Y values set the length and breadth of the Sampled/​Highlighted Video area.  Range is 0 to 100, with the value of 0 covering 0% of the video frame and 100 covering 90% of video frame. ​
 +
 +The Z value enables or disables the 200 µm safety limit described in the AUTOFOCUS OPERATION section on page 4.  Setting safety limit enable = 1 enables the safety limit; safety limit enable = 0 disables the safety feature. ​ The default value is 1.  ​
 +
 +<WRAP center round important 60%>
 +Caution: ​ Disabling the safety limit could result in damage to your optics, your sample, or your focus drive.
 +</​WRAP>​
 +
 +<​asi>​$ AL X=80 Y=50 Z=1
 +:​A<​CR><​LF></​asi>​
 +<​asi>​$ AL
 +:N-3 </​asi>​
 +Error indicates missing arguments
 +<​asi>​$ AL X=1000 Y=-12
 +:​N-4</​asi>​
 +Error indicates arguments out of range
 +<​asi>​$ AL X=90 Y=90
 +:N-5 </​asi>​
 +Error indicates operation failed, try entering one argument at a time
 +<​asi>​$ AL X? Y? Z?
 +:A X=80 Y=50 Z=1</​asi>​
 +
 +
 +<--
 +
 +
 +
 +