User Tools

Site Tools


commands:ensync

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
commands:ensync [2016/09/07 22:03]
asiadmin ↷ Page moved from documentation:commands:ensync to commands:ensync
commands:ensync [2019/04/18 23:33] (current)
Line 1: Line 1:
 +====== Command:​ENSYNC (ES) ======
  
 +--> MS2000 or RM2000 syntax#
 +
 +^Shortcut ​ |ES   ​| ​
 +^Format ​ |ENSYNC [axis]=[position in mm]… |
 +^Units ​   |position in millimeter|
 +^Remembered |Using SS Z |
 +
 +<--
 +
 +--> Tiger syntax# ​
 +
 +^Shortcut ​ |ES   ​| ​
 +^Format ​ |ENSYNC [axis]=[position in mm]… |
 +^Units ​   |position in millimeter|
 +^Type    |Axis-Specific ​ |
 +^Remembered |Using [addr#]SS Z |
 +
 +<--
 +
 +This command lets the user set a position, in millimeters - absolute, which will toggle a TTL output when the stage crosses that position. When ''​ENSYNC''​ is issued, the TTL output is reset low.  Whenever the stage crosses the ''​ENSYNC''​ position, the output will toggle low to high and if crossed again, from high to low.  ''​ENSYNC''​ will only work with one axis at a time, either X or Y and depends on how JP1 is jumped. ​
 +
 +== On MS2000 ==
 +(JP1-1&​2 = X axis, JP1 – 2&3 = Y axis)  The TTL output is available on pin SV1-7. ​ Contact ASI for additional details on these modifications
 +
 +== On Tiger ==
 +Need Dual Axis card Rev F3 and above. Pin is routed to the backplane. Additional hardware is needed to expose the signal. Contact ASI for additional details on these modifications
 +
 +
 +<WRAP center round important 60%>
 +Warning—units of the position info is millimeters rather than tenths of microns
 +</​WRAP>​
 +
 +
 +
 +
 +{{tag>​commands tiger ms2000}}