Line 27: Line 27:
 **Z** ''[lock_range]'': The Z parameter controls the maximum excursion (in either direction) of the stage from the position where the Lock state was initiated before the system generates an error condition and unlocks.  The value ''lock_range'' is in units of millimeters.  The default value is 1.0 mm.   **Z** ''[lock_range]'': The Z parameter controls the maximum excursion (in either direction) of the stage from the position where the Lock state was initiated before the system generates an error condition and unlocks.  The value ''lock_range'' is in units of millimeters.  The default value is 1.0 mm.  
  
-**F** ''[cal_range]'': The F parameter controls the excursion of the stage when going through the calibration sequence.  The default value for ''cal_range'' is 0.005 mm.  Setting the objective's NA using ''LR Y'' will override this value.+**F** ''[cal_range]'': The F parameter controls the excursion of the stage in the dither state of the calibration sequence.  The default value for ''cal_range'' is 0.005 mm.  Setting the objective's NA using ''LR Y'' will change this value.
  
 **T** ''[loop_gain]'': The T parameter controls the gain multiplier or loop gain.  The default value is 4. **T** ''[loop_gain]'': The T parameter controls the gain multiplier or loop gain.  The default value is 4.
Address: 29391 W Enid Rd. Eugene, OR 97402, USA | Phone: +1 (541) 461-8181
commands/lockrg.txt · Last modified: 2024/02/05 20:08 by jon