 +====== Command:PZC ======
  
 +--> MS2000 or RM2000 syntax#
 +
 +^Format ​ |PZC X=[0 or 1] Y=[0,1,2,3] Z=[1 to 100] F=[1 to 100] ,or \\ PZC|
 +^Units ​   |integer codes|
 +^Remembered |Using SS Z |
 +
 +<--
 +
 +--> Tiger syntax# ​
 +
 +^Format ​ |[addr#]PZC X=[0 or 1] Y=[0,1,2,3] Z=[1 to 100] F=[1 to 100] ,or \\ [addr#]PZC|
 +^Units ​   |integer codes|
 +^Type    |Card-Addressed|
 +^Remembered |Using [addr#]SS Z |
 +
 +<--
 +
 +PZC when entered alone runs the auto calibration routine that sets various internal parameters for optimal operation of the piezo top-plate. <wrap reply> :A </​wrap>​ is returned on completion, <wrap reply> :N-5 </​wrap> ​ is returned if the routine failed. ​
 +
 +**X** argument sets the auto calibration type to perform. 0 is for short calibration (default) i.e. only strain gauges offset is adjusted. While 1 is long calibration routine, with adjusts both strain gauge offset and the feedback gain. You will need a length gauge to run the full calibration routine. [[commands:​saveset|Ss z command]] is not applicable, settings will  revert ​ back to default when controller restarts. Note: Long calibration is not implemented for ADEPT card with TG-1000. Usage will end in an error.
 +
 +**Y** argument sets the axis index to which the length gauge is assigned. Default is 0 i.e. X index in a 4 axis build. [[commands:​saveset|Ss z command]] not applicable, settings will  revert ​ back to default on controller restart. ​
 +
 +**Z** argument sets the delay between routine runs, default is 35 i.e. 35ms. Units are in milliseconds. [[commands:​saveset|Ss z command]] not applicable, settings will  revert ​ back to default on controller restart. ​
 +
 +**F** argument sets the position where controller moves the piezo top-plate before adjusting the strain gauge offset. Accepts values between 1 to 100, units are %, default is 50 i.e. middle of the piezo range. [[commands:​saveset|Ss z command]] is applicable, settings will  be saved between controller restarts.  ​
 +
 +Please use [[commands:​halt|HALT command]] to stop a running calibration routine; else the routine will leave incorrect settings on the ADEPT card. 
 +
 +
 +
 +