 +====== Command:SAO ======
 +
 +--> MS2000 or RM2000 syntax#
 +
 +^Shortcut ​ |SAO| 
 +^Format ​ |SAO [axis]=### … |
 +^Units ​   |Axis units|
 +^Remembered |Using SS Z |
 +
 +<--
 +
 +--> Tiger syntax# ​
 +
 +^Shortcut ​ |SAO| 
 +^Format ​ |SAO [axis]=### … |
 +^Units ​   |Axis units|
 +^Type    |Axis-Specific ​ |
 +^Remembered |Using [addr#]SS Z |
 +
 +<--
 +
 +This command sets the position of the center position of the single-axis pattern. ​ For example, if the offset is 1000 and the peak-to-peak amplitude is set to 1000, the pattern will go between positions 500 and 1500.
 +
 +Note that manual moves (e.g. with joystick or wheels) while a single-axis pattern is being generated automatically adjust the single-axis offset value, such that this command may not be needed.
 +
 +As of Tiger firmware v2.82, using the “+” operator instead of specifying a position will store the current position to the offset for the specified axis.  For example, ''​SAO A+''​ set the position of axis A to the center position of the single-axis pattern
 +
 +== Example ==
 +
 +<asi> $ SAO A=1000  ​
 +:​A</​asi>​
 +Sets offset of the offset of the pattern to be 1 degree for (MicroMirror) axis A.
 +
 +
 +
 +
 +
 +