 +====== Command:​SAVESET (SS) ======
  
 +--> MS2000 or RM2000 syntax#
 +
 +^Shortcut ​ |SS| 
 +^Format ​ |SAVESET [X][Y][Z] |
 +
 +<--
 +
 +--> Tiger syntax# ​
 +
 +^Shortcut ​ |SS| 
 +^Format ​ |[addr#​]SAVESET [X][Y][Z] |
 +^Type    |Card-Addressed ​ |
 +
 +<--
 +
 +''​SAVESET''​ allows the user to save current parameters settings to Flash memory. \\
 +''​SAVESET Z''​ , saves settings to flash memory \\
 +''​SAVESET Y''​ , restores previously saved settings after a ''​SAVESET X''​ \\
 +''​SAVESET X''​ , will reload factory defaults upon next power-up \\
 +
 +==Reply==
 +Upon the start of execution of this command, the controller will reply with a  <wrap reply> : </​wrap>​.
 +
 +When the execution is complete, an  <wrap reply> A </​wrap>​ will follow the colon.
 +
 +**Note 1:**  During the time interval between the <wrap reply> : </​wrap>​ and the <wrap reply> A </​wrap>,​ no serial or manual moves should be given.
 +
 +**Note 2:**  In MS2000 Versions 6.1u and later (see ''​VERSION''​ command), limit settings (see ''​SETLOW''​ and ''​SETUP''​) are saved if and only if the ''​SAVEPOS''​ command is issued after the command ''​SAVESET Z''​.
 +
 +
 +== Example ==
 +
 +--> MS2000 example#
 +
 +<asi> $ SS Z
 +:​A</​asi>​
 +Saves current settings to flash memory.
 +<--
 +
 +--> Tiger example#
 +
 +<asi> $ 1SS Z
 +:​A</​asi>​
 +Saves Card#1 's current settings to flash memory.
 +<--
 +
 +{{tag>​commands tiger ms2000 sequencer}}