User Tools

Site Tools


commands:unlock

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
commands:unlock [2019/08/22 17:11]
vik
commands:unlock [2019/08/22 17:12] (current)
vik
Line 24: Line 24:
 Rel_LK_knob_spd (default ​ 2) controls the sensitivity of the control focus knob when the system is locked. ​ This will vary depending on the calibration factor that the system finds, so don’t be alarmed if you find large sensitivity differences with conditions. Rel_LK_knob_spd (default ​ 2) controls the sensitivity of the control focus knob when the system is locked. ​ This will vary depending on the calibration factor that the system finds, so don’t be alarmed if you find large sensitivity differences with conditions.
  
-//__To select active Z-focus axis:__// If the controller can handle more than one Z-axis focus device, you can specify the focus_index to select which one is active for the CRISP, TRACKING or ZS functions. ​ Specify as a 0-indexed number; find the axis index from the letter using Z2B command or else parse the output of the BU X command and figure out what number in order the axis letter is).  ​Save ​the parameter change (SS Z) and reset the controller ​for setting to take effect.  ​+//__To select active Z-focus axis:__// If the controller can handle more than one Z-axis focus device, you can specify the focus_index to select which one is active for the CRISP, TRACKING or ZS functions. ​ Specify as a 0-indexed number; find the axis index from the letter using Z2B command or else parse the output of the BU X command and figure out what number in order the axis letter is). S__ave ​the parameter change ([[commands:​saveset|SS Z]]) and reset the controller__ ​for setting to take effect.  ​
  
 <-- <--
Address: 29391 W Enid Rd. Eugene, OR 97402, USA | Phone: +1 (541) 461-8181
commands/unlock.txt · Last modified: 2019/08/22 17:12 by vik