epi_illumination_options [2016/10/04 15:13]
127.0.0.1 external edit
epi_illumination_options [2018/11/09 12:33] (current)
jon [Kohler Illumination]
Line 49: Line 49:
  
 There are two constraints that must both be satisfied to get optimum results.  ​ There are two constraints that must both be satisfied to get optimum results.  ​
-  - Ordered List ItemCollimated ​light from the illuminator needs to be refocused at approximately the objective lens back focal plane.  +  - Collimated ​light from the illuminator needs to be refocused at approximately the objective lens back focal plane. This constraint is generally not met exactly unless care is taken, but ASI's LED sources can be easily adjusted in the field to meet this condition
-  - The illuminator iris must be at the focal point of the tube lens.  This second constraint ​is guaranteed by the construction conventions of the ASI tube lenses and the length of the ''​C60-LAMP_ADPT''​.  However, the first constraint is generally not met exactly unless care is taken to be sure that it is.+  - The illuminator iris must be at the focal point of the tube lens.  This is guaranteed by the construction conventions of the ASI tube lenses and the length of the ''​C60-LAMP_ADPT''​.
  
 ==== Focusing to objective back focal plane ==== ==== Focusing to objective back focal plane ====