 **Note: the terms left and right refer to the sides of the microscope as viewed from the front.** **Note: the terms left and right refer to the sides of the microscope as viewed from the front.**
-[{{ :​documentation:​aristoplan_1.png?​direct&​300 |Figure 1.}}]+[{{ aristoplan_1.png?​direct&​300 |Figure 1.}}]
  
-[{{ :​documentation:​aristoplan_2.png?​direct&​300 |Figure 2.}}]  +[{{ aristoplan_2.png?​direct&​300 |Figure 2.}}]  
  
 1) Remove the left fine focus knob by using the 1.5mm hex wrench to loosen the set screw as shown     above in figure 1.  Then pull the knob off of the fine focus shaft as shown in figure 2.  1) Remove the left fine focus knob by using the 1.5mm hex wrench to loosen the set screw as shown     above in figure 1.  Then pull the knob off of the fine focus shaft as shown in figure 2. 
Line 26: Line 26:
 Locate the motor drive and the back plate. ​ If the drive is not installed onto the back plate, install it as shown in figure 3.  Do not completely tighten the horizontal adjustment screw shown in figure 3.  Leave it loose enough so that the drive can easily slide within the grove. ​ Also loosen the vertical adjustment screws shown in figure 4 so that the drive can easily slide within the grove. ​ At this time, also loosen the screw on the fine focus shaft clamp that is located near the black encoder. ​ This clamp is showed installed in figure 5.  Locate the motor drive and the back plate. ​ If the drive is not installed onto the back plate, install it as shown in figure 3.  Do not completely tighten the horizontal adjustment screw shown in figure 3.  Leave it loose enough so that the drive can easily slide within the grove. ​ Also loosen the vertical adjustment screws shown in figure 4 so that the drive can easily slide within the grove. ​ At this time, also loosen the screw on the fine focus shaft clamp that is located near the black encoder. ​ This clamp is showed installed in figure 5. 
  
-[{{ :​documentation:​aristoplan_3.png?​direct&​300 |Figure 3}}]   ​  ​    +[{{ aristoplan_3.png?​direct&​300 |Figure 3}}]   ​  ​    
  
-[{{ :​documentation:​aristoplan_4.png?​direct&​300 |Figure 4}}]+[{{ aristoplan_4.png?​direct&​300 |Figure 4}}]
  
-[{{ :​documentation:​aristoplan_5.png?​direct&​300 |Figure 5}}]+[{{ aristoplan_5.png?​direct&​300 |Figure 5}}]
  
-[{{ :​documentation:​aristoplan_6.png?​direct&​300 |Figure 6}}]+[{{ aristoplan_6.png?​direct&​300 |Figure 6}}]
  
-[{{ :​documentation:​aristoplan_7.png?​direct&​300 |Figure 7}}]+[{{ aristoplan_7.png?​direct&​300 |Figure 7}}]
  
-[{{ :​documentation:​aristoplan_8.png?​direct&​300 |Figure 8}}]+[{{ aristoplan_8.png?​direct&​300 |Figure 8}}]
  
 In this step we will install the ASI backplate and motor drive assembly as a unit as shown in figure 6. The ASI back plate will be installed and clamped to the back of the microscope as shown in figures 7 & 8 only after the motor drive is clamped onto the fine focus shaft of the microscope as shown in figure 5.  ​ In this step we will install the ASI backplate and motor drive assembly as a unit as shown in figure 6. The ASI back plate will be installed and clamped to the back of the microscope as shown in figures 7 & 8 only after the motor drive is clamped onto the fine focus shaft of the microscope as shown in figure 5.  ​
Line 47: Line 47:
   - Tighten the vertical and horizontal adjustment screws in their mid positions as shown in figures 9 & 10.   - Tighten the vertical and horizontal adjustment screws in their mid positions as shown in figures 9 & 10.
  
-[{{ :​documentation:​aristoplan_9.png?​direct&​300 |Figure 9 }}]+[{{ aristoplan_9.png?​direct&​300 |Figure 9 }}]
             
-[{{ :​documentation:​aristoplan_10.png?​direct&​300 |Figure 10: Position back plate so that vertical adjustment screws are midway in slots as shown in figure 9 & 10}}]+[{{ aristoplan_10.png?​direct&​300 |Figure 10: Position back plate so that vertical adjustment screws are midway in slots as shown in figure 9 & 10}}]
  
-[{{ :​documentation:​aristoplan_11.png?​direct&​300 |Figure 11: Use the 5/64” Allen Wrench to tighten the +[{{ aristoplan_11.png?​direct&​300 |Figure 11: Use the 5/64” Allen Wrench to tighten the 
 Base plate clamp at a point where the vertical adjustment screws are midway in the slots  Base plate clamp at a point where the vertical adjustment screws are midway in the slots 
 }}] }}]
Line 62: Line 62:
 ==== Step 6 - Installing the cover & left fine focus knob ==== ==== Step 6 - Installing the cover & left fine focus knob ====
  
-[{{ :​documentation:​aristoplan_12.png?​direct&​300 |Figure 12 }}]+[{{ aristoplan_12.png?​direct&​300 |Figure 12 }}]
  
-[{{ :​documentation:​aristoplan_13.png?​direct&​300 |Figure 13}}]   ​+[{{ aristoplan_13.png?​direct&​300 |Figure 13}}]   ​
  
-[{{ :​documentation:​aristoplan_14.png?​direct&​300 |Figure 14 }}]+[{{ aristoplan_14.png?​direct&​300 |Figure 14 }}]
  
-[{{ :​documentation:​aristoplan_15.png?​direct&​300 |Figure 15}}]+[{{ aristoplan_15.png?​direct&​300 |Figure 15}}]
  
  