 ===== Part 1 - Removing the Left Fine Focus Knob ===== ===== Part 1 - Removing the Left Fine Focus Knob =====
    
-[{{ :​documentation:​dmirb_1.png?​direct&​300 |Figure 1. Left Fine Focus Knob Removed}}] ​+[{{ dmirb_1.png?​direct&​300 |Figure 1. Left Fine Focus Knob Removed}}] ​
  
 Remove the left fine focus knob from the microscope as follows:  ​ Remove the left fine focus knob from the microscope as follows:  ​
Line 27: Line 27:
 ===== Part-2 Installing & Aligning the Motor Drive ===== ===== Part-2 Installing & Aligning the Motor Drive =====
  
-[{{ :​documentation:​dmirb_2.png?​direct&​300 |Figure 2. ASI Drive Installed on Leica DMIRB Microscope}}] ​+[{{ dmirb_2.png?​direct&​300 |Figure 2. ASI Drive Installed on Leica DMIRB Microscope}}] ​
  
  
Line 36: Line 36:
 The ASI motor drive attaches to the silver horizontal adjustment bar that is attached to the base plate. The motor drive slides within two lips on either side of the horizontal adjustment bar so that it can slide along the vertical axis and is secured to the adjustment bar via two 4/40 vertical adjustment screws. The adjustment bar is secured to the baseplate with a 4x12mm horizontal adjustment screw. The horizontal and vertical adjustments allow the drive to be correctly positioned so the drive shaft on the ASI motor drive can be slid over the fine focus shaft of the microscope.  ​ The ASI motor drive attaches to the silver horizontal adjustment bar that is attached to the base plate. The motor drive slides within two lips on either side of the horizontal adjustment bar so that it can slide along the vertical axis and is secured to the adjustment bar via two 4/40 vertical adjustment screws. The adjustment bar is secured to the baseplate with a 4x12mm horizontal adjustment screw. The horizontal and vertical adjustments allow the drive to be correctly positioned so the drive shaft on the ASI motor drive can be slid over the fine focus shaft of the microscope.  ​
  
-[{{ :​documentation:​dmirb_3.png?​direct&​300 |Figure 3. The Horizontal Adjustment Screw}}]+[{{ dmirb_3.png?​direct&​300 |Figure 3. The Horizontal Adjustment Screw}}]
  
 a) Locate the base plate, the motor drive, adjustment bar, and the Allen wrenches. ​ Use the 3mm Allen wrench to remove the drive assembly from the base plate by removing the horizontal adjustment screw as shown in figure 3.  a) Locate the base plate, the motor drive, adjustment bar, and the Allen wrenches. ​ Use the 3mm Allen wrench to remove the drive assembly from the base plate by removing the horizontal adjustment screw as shown in figure 3. 
  
-[{{ :​documentation:​dmirb_4.png?​direct&​300 | Figure 4. Base plate should be flush with the side of the microscope and about center under the focus knob.}}]+[{{ dmirb_4.png?​direct&​300 | Figure 4. Base plate should be flush with the side of the microscope and about center under the focus knob.}}]
  
-[{{ :​documentation:​dmirb_5.png?​direct&​300 |Figure 5. Clamp to tighten base plate to  the microscope.}}] ​                                    +[{{ dmirb_5.png?​direct&​300 |Figure 5. Clamp to tighten base plate to  the microscope.}}] ​                                    
  
 b)  Slide the baseplate assembly under the microscope until the baseplate is about centered under the focus knob. Insure that the base plate is parallel with the bottom of the microscope. Position the base plate so that it is under the focus assembly as shown in figure 4. b)  Slide the baseplate assembly under the microscope until the baseplate is about centered under the focus knob. Insure that the base plate is parallel with the bottom of the microscope. Position the base plate so that it is under the focus assembly as shown in figure 4.
  
-[{{ :​documentation:​dmirb_6.png?​direct&​300 |Figure 6. The vertical adjustment screws}}]+[{{ dmirb_6.png?​direct&​300 |Figure 6. The vertical adjustment screws}}]
  
-[{{ :​documentation:​dmirb_7.png?​direct&​300 |Figure 7. The drive shaft clamp}}]  +[{{ dmirb_7.png?​direct&​300 |Figure 7. The drive shaft clamp}}]  
  
 c) Locate the adjustment bar, and the motor drive. ​ Use the 7/64" Allen wrench to loosen the vertical adjustment screws as shown in figure 6. Use the 7/64 Allen wrench to loosen the clamp that is located on the ASI drive shaft as shown in  Figure 7. This clamp is located at the end of the drive shaft as it protrudes out of the  black encoder cover. ​ Once the clamp is loose slide it back towards the encoder. While holding the right fine focus knob with one hand align the ASI drive shaft with the microscope’s fine focus shaft and slide the ASI drive shaft over the microscope’s fine focus shaft. ​ If you do not hold the right fine focus knob the microscope’s fine focus shaft may be pressed over towards the right of the microscope. ​ If this happens simply push the right fine focus knob in towards the microscope to push the shaft back.    c) Locate the adjustment bar, and the motor drive. ​ Use the 7/64" Allen wrench to loosen the vertical adjustment screws as shown in figure 6. Use the 7/64 Allen wrench to loosen the clamp that is located on the ASI drive shaft as shown in  Figure 7. This clamp is located at the end of the drive shaft as it protrudes out of the  black encoder cover. ​ Once the clamp is loose slide it back towards the encoder. While holding the right fine focus knob with one hand align the ASI drive shaft with the microscope’s fine focus shaft and slide the ASI drive shaft over the microscope’s fine focus shaft. ​ If you do not hold the right fine focus knob the microscope’s fine focus shaft may be pressed over towards the right of the microscope. ​ If this happens simply push the right fine focus knob in towards the microscope to push the shaft back.   
Line 62: Line 62:
  
  
-[{{ :​documentation:​dmirb_8.png?​direct&​300 |Figure 8. Installing the cover & the fine focus knob}}]+[{{ dmirb_8.png?​direct&​300 |Figure 8. Installing the cover & the fine focus knob}}]
  
 ===== Part 3-Installing the motor drive cover plate & fine focus knob ===== ===== Part 3-Installing the motor drive cover plate & fine focus knob =====