+ | ====== Installation of ASI MS-2000 Automated XY stage onto Leica DMRB Microscope ====== | ||
+ | This procedure outlines the steps necessary to install the ASI MS-2000 automated XY stage onto Leica DMRB microscope. | ||
+ | |||
+ | - Removing the condenser carrier. | ||
+ | - Removing the original mechanical xy stage. | ||
+ | - Installing the ASI MS-2000 automated XY stage. | ||
+ | - Replacing the condenser carrier. | ||
+ | |||
+ | ===== Step # 1 Removing the condenser carrier. ===== | ||
+ | |||
+ | |||
+ | To gain access to the 3 ea. 4x8mm screws that secure the original mechanical xy stage the condenser carrier must be removed. | ||
+ | |||
+ | |||
+ | [{{ dmrb_leica_2_.jpg? | ||
+ | |||
+ | |||
+ | ===== Step 2 Removing the original mechanical xy stage. ===== | ||
+ | |||
+ | |||
+ | Once the condenser carrier has been removed then the original mechanical stage can be taken off the microscope. | ||
+ | |||
+ | [{{ dmrb_leica_5.png? | ||
+ | |||
+ | [{{ dmrb_leica_4_.jpg? | ||
+ | |||
+ | ===== Step 3 Installing the ASI MS-2000 automated XY stage ===== | ||
+ | |||
+ | Locate the ASI stage and align it as shown in figure 3 so that the holes in the stage align with the holes in the stage carrier. | ||
+ | |||
+ | [{{ dmrb_leica_1_.jpg? | ||
+ | |||
+ | ===== Step 4 Replacing the condenser carrier ===== | ||
+ | |||
+ | |||
+ | Replace the condenser carrier in the reverse order as given in [[leica_dmrb_xystage# | ||
+ | |||