+ | ====== Installing the MS-2000 XY Stage onto your Inverted Microscope ====== | ||
+ | |||
+ | ===== The procedure has three parts: ===== | ||
+ | |||
+ | Removing the original stage. \\ | ||
+ | Install the MS-2000 XY stage on the microscope \\ | ||
+ | Connecting the cables on the MS-2000 control unit \\ | ||
+ | |||
+ | ===== Part 1 – Remove the Original Stage ===== | ||
+ | |||
+ | |||
+ | Before removing or installing the stage, the objectives should be removed from the nosepiece and stored in a safe and secure place until after the stage has been installed and aligned. | ||
+ | |||
+ | ===== Part 2 – Install the MS-2000 XY Stage on the Microscope ===== | ||
+ | |||
+ | |||
+ | Remove any protective stage shipping bolts or brackets. Mount the stage with the new screws supplied with the ASI stage. | ||
+ | On some flat top stages, (as seen in the inset photos at right), the screw holes are not visible until the stage is moved under power to the left and right on the Y-axis. This requires that __part 3__ be completed first. | ||
+ | |||
+ | {{ ms2000_stage_install1.png? | ||
+ | |||
+ | |||
+ | {{ ms2000_stage_install2.png? | ||
+ | |||
+ | ===== Part 3 – Connecting the Cables on the MS-2000 Control Unit ===== | ||
+ | |||
+ | |||
+ | The XY stage’s 5 ft. cable connects to the controller’s 25-pin female DB-Connector. | ||
+ | |||
+ | |||
+ | {{tag> |