+ | ====== Nikon Eclipse 600 Installation Procedure ====== | ||
+ | The procedure below outlines the steps necessary to install the ASI Microscope Focus Controller Drive onto the Nikon Eclipse | ||
+ | |||
+ | To perform the following steps you will need the following tools: | ||
+ | |||
+ | Medium flat blade screw driver | ||
+ | 3mm, 5/64 and 7/64 inch hex wrenches | ||
+ | The hex wrenches are provided by ASI. | ||
+ | |||
+ | The procedure has three parts: | ||
+ | |||
+ | 1) Removing the left fine focus knob \\ | ||
+ | 2) Installing the baseplate. \\ | ||
+ | 3) Installing and aligning the motor drive assembly. \\ | ||
+ | 4) Installing the motor drive cover plate & fine focus knob. \\ | ||
+ | |||
+ | ===== Part 1 - Removing the Left Fine Focus Knob ===== | ||
+ | |||
+ | Remove the left fine focus knob from the microscope as follows: | ||
+ | |||
+ | [{{ e600_1.png? | ||
+ | |||
+ | [{{ e600_2.png? | ||
+ | |||
+ | [{{ e600_3.png? | ||
+ | |||
+ | a) Pull off the black rubber boot covering the fine focus knob. \\ | ||
+ | b) Remove the round slotted nut from the end of the knob using the medium slotted screwdriver. \\ | ||
+ | c) Remove the fine focus knob by grasping it firmly and pulling it straight off. Do not turn the knob to work it off as this will damage the plastic alignment key molded into the knob. \\ | ||
+ | |||
+ | ===== Part 2 - Installing the Baseplate ===== | ||
+ | |||
+ | [{{ e600_4.png? | ||
+ | |||
+ | [{{ e600_5.png? | ||
+ | | ||
+ | Locate the baseplate and position it so that the horizontal adjustment bar is on the left side of the microscope. Lift the front of the microscope up and slide the baseplate under the microscope so that the horizontal adjustment bar is centered under the left focus knob. | ||
+ | |||
+ | |||
+ | ===== Part-3 Installing & Aligning the Motor Drive ===== | ||
+ | |||
+ | [{{ e600_6.png? | ||
+ | |||
+ | [{{ e600_7.png? | ||
+ | |||
+ | [{{ e600_8.png? | ||
+ | |||
+ | a) Loosen the screw securing the adjustment bar to the baseplate. | ||
+ | |||
+ | b) Locate the ASI fine focus motor drive and vertical adjustment screw (M4 x 12 hex head screw).Align the drive shaft on the ASI drive with the microscope fine focus shaft and slide the motor drive over the fine focus shaft. | ||
+ | |||
+ | c) Rotate the drive plate so that the recesses at the bottom, inside edges of the drive plate align with the keys on the outside edges of the adjustment bar, then press the drive plate towards the adjustment bar until it meshes. | ||
+ | |||
+ | [{{ e600_9.png? | ||
+ | |||
+ | d) Screw in the vertical adjustment screw but leave it just loose enough so that the drive plate can slide up and down. Use the 3 mm hex wrench. | ||
+ | |||
+ | e) Slide the motor drive up and down, forward and backward slightly while turning the right fine focus knob until it is in the position where minimum drag is felt on the right focus knob. Secure the motor drive into position by tightening the horizontal and vertical adjustment screws. Slide the tab that protrudes from the ASI drive shaft clamp into one of the slots on the white plastic gear on the microscope focus shaft, as shown in figure # 6. This clamp is located on the drive shaft between the black encoder and the microscope. | ||
+ | |||
+ | [{{ e600_10.png? | ||
+ | |||
+ | g) Recheck the alignment by noting the drag on the right fine focus knob. No noticeable drag should be felt Repeat the steps above if necessary. | ||
+ | |||
+ | |||
+ | ===== Part 4-Installing the motor drive cover plate & fine focus knob. ===== | ||
+ | |||
+ | {{ e600_11.png? | ||
+ | |||
+ | {{ e600_12.png? | ||
+ | |||
+ | a) Locate the motor drive cover. Remove the 4/40 button head screws from the drive with the 1/16 “ Allen wrench. Position the motor drive covert over the motor drive assembly and secure in place using the 4/40 button head screws. | ||
+ | |||
+ | b) Slide the microscope fine focus knob over the shaft extension and press it on all | ||
+ | the way so that the plastic tabs on the knob mate with the slots on the ASI drive shaft. | ||
+ | |||
+ | This completes the procedure for installing the ASI motor drive on to the Nikon Eclipse Microscope. | ||
+ | |||