Line 20: Line 20:
 Remove the right fine focus knob from the microscope as follows:  ​ Remove the right fine focus knob from the microscope as follows:  ​
  
-[{{ :​documentation:​600fn_1.png?​direct&​300 |Figure 1: Use screwdriver to remove nut}}]+[{{ 600fn_1.png?​direct&​300 |Figure 1: Use screwdriver to remove nut}}]
  
-[{{ :​documentation:​600fn_2.png?​direct&​300 |Figure 2: Pull knob straight off to expose shaft}}] ​    +[{{ 600fn_2.png?​direct&​300 |Figure 2: Pull knob straight off to expose shaft}}] ​    
  
  
Line 33: Line 33:
    
  
-[{{ :​documentation:​600fn_3.png?​direct&​150 |Figure 4a Installing ​ baseplate }}]+[{{ 600fn_3.png?​direct&​150 |Figure 4a Installing ​ baseplate }}]
                                  
-[{{ :​documentation:​600fn_4.png?​direct&​300 |Figure 4b Installed baseplate }}]+[{{ 600fn_4.png?​direct&​300 |Figure 4b Installed baseplate }}]
   ​         ​   ​         ​
-[{{ :​documentation:​600fn_5.png?​direct&​300 |Figure 4c On some early drives two washers were used to position the drive out. }}]+[{{ 600fn_5.png?​direct&​300 |Figure 4c On some early drives two washers were used to position the drive out. }}]
  
 Locate the ASI drive and baseplate assembly. ​ If the drive is attached to the base plate use the Allen wrench provided to remove the horizontal adjustment screw to remove the drive. ​ Use the screws and Allen wrench provided to install the baseplate as outlined in figures 4a - 4c above.         Locate the ASI drive and baseplate assembly. ​ If the drive is attached to the base plate use the Allen wrench provided to remove the horizontal adjustment screw to remove the drive. ​ Use the screws and Allen wrench provided to install the baseplate as outlined in figures 4a - 4c above.        
Line 43: Line 43:
 ===== Part-3 Installing & Aligning the Motor Drive ===== ===== Part-3 Installing & Aligning the Motor Drive =====
  
-[{{ :​documentation:​600fn_6.png?​direct&​300 |Figure 5:  Slide ASI dive onto microscope'​s fine focus shaft}}]+[{{ 600fn_6.png?​direct&​300 |Figure 5:  Slide ASI dive onto microscope'​s fine focus shaft}}]
  
-[{{ :​documentation:​600fn_7.png?​direct&​300 |Figure 6: Tab on adjustment bar slides into groove on baseplate}}] ​                  +[{{ 600fn_7.png?​direct&​300 |Figure 6: Tab on adjustment bar slides into groove on baseplate}}] ​                  
  
-[{{ :​documentation:​600fn_8.png?​direct&​300 |Figure 7: Secure, but Do Not tighten , horizontal adjustment screw  to hold drive in }}]                           +[{{ 600fn_8.png?​direct&​300 |Figure 7: Secure, but Do Not tighten , horizontal adjustment screw  to hold drive in }}]                           
  
-[{{ :​documentation:​600fn_9.png?​direct&​300 |Figure 8: Vertical adjustment screws should still be loose. }}]             ​+[{{ 600fn_9.png?​direct&​300 |Figure 8: Vertical adjustment screws should still be loose. }}]             ​
  
-[{{ :​documentation:​600fn_10.png?​direct&​300 |Figure 9: Metal tab on drive shaft clamp slides into slot on plastic gear}}]+[{{ 600fn_10.png?​direct&​300 |Figure 9: Metal tab on drive shaft clamp slides into slot on plastic gear}}]
  ​  ​          ​  ​        
-[{{ :​documentation:​600fn_11.png?​direct&​300 |Figure 10: While holding drive shaft clamp in place securely tighten the drive shaft clamp. ​ This clamp must be securely tightened or slippage may occur. }}]+[{{ 600fn_11.png?​direct&​300 |Figure 10: While holding drive shaft clamp in place securely tighten the drive shaft clamp. ​ This clamp must be securely tightened or slippage may occur. }}]
  
 As shown in figures 5-10 above, install the ASI z-axis drive onto the microscope'​s fine focus shaft. ​ Once you have securely tighten the drive shaft clamp as shown in figure 10 rotate the left fine focus knob.  No drag should be felt.  If no drag is felt tighten the Horizontal and Vertical adjustment screws as shown in figures 7 & 8.   ​Recheck the alignment by turning the left fine focus knob.  No noticeable drag should be felt.  If drag is felt at any one point with in a full rotation loosen the Horizontal and Vertical adjustment screws. Then slide the motor drive up and down, forward and backward slightly while turning the left fine focus knob until it is in the position where minimum drag is felt.  Secure the motor drive into position by tightening the horizontal and vertical adjustment screws at this point. ​ Since the drive is usually self aligning this second set is usually not required. As shown in figures 5-10 above, install the ASI z-axis drive onto the microscope'​s fine focus shaft. ​ Once you have securely tighten the drive shaft clamp as shown in figure 10 rotate the left fine focus knob.  No drag should be felt.  If no drag is felt tighten the Horizontal and Vertical adjustment screws as shown in figures 7 & 8.   ​Recheck the alignment by turning the left fine focus knob.  No noticeable drag should be felt.  If drag is felt at any one point with in a full rotation loosen the Horizontal and Vertical adjustment screws. Then slide the motor drive up and down, forward and backward slightly while turning the left fine focus knob until it is in the position where minimum drag is felt.  Secure the motor drive into position by tightening the horizontal and vertical adjustment screws at this point. ​ Since the drive is usually self aligning this second set is usually not required.
Line 60: Line 60:
  
  
-[{{ :​documentation:​600fn_12.png?​direct&​300 |Figure 11}}]+[{{ 600fn_12.png?​direct&​300 |Figure 11}}]
                  ​                  ​
 Locate the cover and install it on the drive as shown using the Allen wrench and the supplied screws that are in the drive plate. ​ Before tightening the screws make sure the cover is positioned so that it does not rub against the outer clamp as shown. ​ Once the cover has been installed the original knob can be installed onto the protruding shaft. ​ Note that the slots on the shaft mate with the plastic tabs on the knob. Locate the cover and install it on the drive as shown using the Allen wrench and the supplied screws that are in the drive plate. ​ Before tightening the screws make sure the cover is positioned so that it does not rub against the outer clamp as shown. ​ Once the cover has been installed the original knob can be installed onto the protruding shaft. ​ Note that the slots on the shaft mate with the plastic tabs on the knob.