 This procedure steps you through the installation and alignment of the ASI  Z-axis motor drive onto the Nikon Eclipse Ni microscope. This procedure steps you through the installation and alignment of the ASI  Z-axis motor drive onto the Nikon Eclipse Ni microscope.
  
-{{ :​documentation:​eni_1.png?​direct&​300 |}} +{{ eni_1.png?​direct&​300 |}} 
  
  
Line 25: Line 25:
 1. Using a flathead screwdriver,​ carefully pry off the fine focus knob off the Left side of the microscope. ​   The knob is held on by a magnetic clamp which should release with a little amount of pressure. ​ 1. Using a flathead screwdriver,​ carefully pry off the fine focus knob off the Left side of the microscope. ​   The knob is held on by a magnetic clamp which should release with a little amount of pressure. ​
        
-[{{ :​documentation:​eni_2.png?​direct&​300 |The terms ‘left’ and ‘right’ refer to the sides of the microscope as viewed from the front of the microscope.}}]+[{{ eni_2.png?​direct&​300 |The terms ‘left’ and ‘right’ refer to the sides of the microscope as viewed from the front of the microscope.}}]
  
-[{{ :​documentation:​eni_3.png?​direct&​300 |Use a flat head screwdriver to pry loose  fine focus knob.}}]+[{{ eni_3.png?​direct&​300 |Use a flat head screwdriver to pry loose  fine focus knob.}}]
  
-[{{ :​documentation:​eni_4.png?​direct&​300 |Carefully pull on the inner fine focus knob and remove it.   }}]+[{{ eni_4.png?​direct&​300 |Carefully pull on the inner fine focus knob and remove it.   }}]
  
  
Line 39: Line 39:
 3. Slide the Drive Gear on to the microscopes fine focus drive plate so that the gear posts engage the slosts on the plate. Note: The drive gear will magnetically lock onto the plate. In some cases the microscopes find focus drive plate will be slightly warped preventing a flat fit – the flex coupler will compensate for this. 3. Slide the Drive Gear on to the microscopes fine focus drive plate so that the gear posts engage the slosts on the plate. Note: The drive gear will magnetically lock onto the plate. In some cases the microscopes find focus drive plate will be slightly warped preventing a flat fit – the flex coupler will compensate for this.
  
-[{{ :​documentation:​eni_5.png?​direct&​300 |Align posts on Drive Gear to match slots on microscope fine focus plate. ​  ​}}] ​              +[{{ eni_5.png?​direct&​300 |Align posts on Drive Gear to match slots on microscope fine focus plate. ​  ​}}] ​              
  
-[{{ :​documentation:​eni_6.png?​direct&​300 |The drive gear will magnetically lock as flush as possible.}}] ​  +[{{ eni_6.png?​direct&​300 |The drive gear will magnetically lock as flush as possible.}}] ​  
  
 ===== Part-3: ​ Installing Flexible Coupling. ===== ===== Part-3: ​ Installing Flexible Coupling. =====
Line 51: Line 51:
 3. Ensure the Drive Gear is still firmly locked to the microscope. ​   ​ 3. Ensure the Drive Gear is still firmly locked to the microscope. ​   ​
  
-[{{ :​documentation:​eni_7.png?​direct&​300 |Line up the coupler with the shaft.}}]  ​+[{{ eni_7.png?​direct&​300 |Line up the coupler with the shaft.}}]  ​
  
-[{{ :​documentation:​eni_8.png?​direct&​300 |Slide flexible coupler over the Drive Gear’s shaft  while holding right hand knob in place.}}]  ​+[{{ eni_8.png?​direct&​300 |Slide flexible coupler over the Drive Gear’s shaft  while holding right hand knob in place.}}]  ​
  
-[{{ :​documentation:​eni_9.png?​direct&​300 |}}]  ​+[{{ eni_9.png?​direct&​300 |}}]  ​
  
 Ensuring the coupler is all the way on the Drive Gear shaft, lightly tighten the flex coupler set screw.  ​ Ensuring the coupler is all the way on the Drive Gear shaft, lightly tighten the flex coupler set screw.  ​
Line 67: Line 67:
 1. If shipped assembled, use the 3 mm hex wrench and disassemble the Z-drive motor assembly from base plate by removing the horizontal adjustment screw. ​ At this time, also loosen the two vertical adjustment screws. 1. If shipped assembled, use the 3 mm hex wrench and disassemble the Z-drive motor assembly from base plate by removing the horizontal adjustment screw. ​ At this time, also loosen the two vertical adjustment screws.
  
-[{{ :​documentation:​eni_10.png?​direct&​300 |Remove motor drive assembly from base plate (if attached) by using the 3mm hex wrench to remove the horizontal adjustment screw.}}]+[{{ eni_10.png?​direct&​300 |Remove motor drive assembly from base plate (if attached) by using the 3mm hex wrench to remove the horizontal adjustment screw.}}]
  
-[{{ :​documentation:​eni_11.png?​direct&​300 |Use the 3mm hex  wrench to loosen the two vertical adjustment screws.}}]+[{{ eni_11.png?​direct&​300 |Use the 3mm hex  wrench to loosen the two vertical adjustment screws.}}]
  
 2. Slide the baseplate assembly completely under the microscope as shown until the small stop plate located on the baseplate is completely against the side of the microscope. ​ This stop plate should be parallel with the bottom of the microscope. ​ Align the plate beneath the focus knobs. 2. Slide the baseplate assembly completely under the microscope as shown until the small stop plate located on the baseplate is completely against the side of the microscope. ​ This stop plate should be parallel with the bottom of the microscope. ​ Align the plate beneath the focus knobs.
  
-[{{ :​documentation:​eni_12.png?​direct&​300 |Slide base plate under microscope and center beneath focus assembly.}}]+[{{ eni_12.png?​direct&​300 |Slide base plate under microscope and center beneath focus assembly.}}]
  
-[{{ :​documentation:​eni_13.png?​direct&​300 |Leave base plate clamp loose at this time.}}]+[{{ eni_13.png?​direct&​300 |Leave base plate clamp loose at this time.}}]
  
 3. Leave base plate clamp slightly loose for later adjustment. 3. Leave base plate clamp slightly loose for later adjustment.
Line 81: Line 81:
 4. Slide the drive shaft of the motor drive assembly into the flexible coupler. 4. Slide the drive shaft of the motor drive assembly into the flexible coupler.
  
-[{{ :​documentation:​eni_14.png?​direct&​300 |Slide the drive shaft into the flexible coupler.}}]+[{{ eni_14.png?​direct&​300 |Slide the drive shaft into the flexible coupler.}}]
  
-[{{ :​documentation:​eni_15.png?​direct&​300 |Coupler accepts about 4 mm of the drive shaft.}}]+[{{ eni_15.png?​direct&​300 |Coupler accepts about 4 mm of the drive shaft.}}]
  
-[{{ :​documentation:​eni_16.png?​direct&​300 |Reinstall the horizontal adjustment screw and tighten loosely for later adjustment.}}]+[{{ eni_16.png?​direct&​300 |Reinstall the horizontal adjustment screw and tighten loosely for later adjustment.}}]
  
 5. Rotate the adjustment bar so that the lip on the bottom of the adjustment bar mates with the groove in the baseplate. ​ Align the holes on the adjustment bar to those on the drive and the baseplate, and screw in the horizontal adjustment screw to hold the drive assembly in place using the 3mm hex wrench. ​ Leave the adjustment screws loose enough so that later the drive can slide up and down and the adjustment bar can slide back and forth within the baseplate. 5. Rotate the adjustment bar so that the lip on the bottom of the adjustment bar mates with the groove in the baseplate. ​ Align the holes on the adjustment bar to those on the drive and the baseplate, and screw in the horizontal adjustment screw to hold the drive assembly in place using the 3mm hex wrench. ​ Leave the adjustment screws loose enough so that later the drive can slide up and down and the adjustment bar can slide back and forth within the baseplate.
Line 91: Line 91:
 6. After verifying the baseplate and motor drive is appropriately positioned under the microscope, snugly tighten the baseplate clamp on the right side of the microscope. ​ Use the 1.5mm hex wrench to tighten the setscrew located in the middle of the clamp assembly as shown. ​ Tightening this setscrew will cause the silver bar to press against the side of the microscope. ​ Ensure that the setscrew is securely tightened to hold the baseplate assembly in place. 6. After verifying the baseplate and motor drive is appropriately positioned under the microscope, snugly tighten the baseplate clamp on the right side of the microscope. ​ Use the 1.5mm hex wrench to tighten the setscrew located in the middle of the clamp assembly as shown. ​ Tightening this setscrew will cause the silver bar to press against the side of the microscope. ​ Ensure that the setscrew is securely tightened to hold the baseplate assembly in place.
  
-[{{ :​documentation:​eni_17.png?​direct&​300 |Tighten the baseplate clamp snugly.}}]+[{{ eni_17.png?​direct&​300 |Tighten the baseplate clamp snugly.}}]
  
-[{{ :​documentation:​eni_18.png?​direct&​300 |Tighten the flexible coupler’s clamp onto the drive shaft. ​  ​Ensure both screws are tight.}}]+[{{ eni_18.png?​direct&​300 |Tighten the flexible coupler’s clamp onto the drive shaft. ​  ​Ensure both screws are tight.}}]
  
 7. Tighten the clamp of the flexible coupler onto the drive shaft of the Z-drive unit using the 2mm hex wrench. 7. Tighten the clamp of the flexible coupler onto the drive shaft of the Z-drive unit using the 2mm hex wrench.
Line 100: Line 100:
 8. Lightly tighten the vertical and horizontal adjustment screws as shown. ​ Since the motor is usually self-aligning,​ this should be a good position for the drive. ​ Check the alignment by noting the drag while rotating the right-hand fine focus knob.  No noticeable drag should be felt for the full 360º rotation of the fine focus knob.  If any drag is felt other than the slight drag of the gears, loosen the vertical and horizontal adjustment screws and move the drive to a point where no drag is felt.  Then tighten the vertical and horizontal adjustment screws. 8. Lightly tighten the vertical and horizontal adjustment screws as shown. ​ Since the motor is usually self-aligning,​ this should be a good position for the drive. ​ Check the alignment by noting the drag while rotating the right-hand fine focus knob.  No noticeable drag should be felt for the full 360º rotation of the fine focus knob.  If any drag is felt other than the slight drag of the gears, loosen the vertical and horizontal adjustment screws and move the drive to a point where no drag is felt.  Then tighten the vertical and horizontal adjustment screws.
  
-[{{ :​documentation:​eni_19.png?​direct&​300 |Tighten the vertical adjustment screws after alignment.}}]+[{{ eni_19.png?​direct&​300 |Tighten the vertical adjustment screws after alignment.}}]
  
-[{{ :​documentation:​eni_20.png?​direct&​300 |Tighten the horizontal adjustment screws after alignment.}}]+[{{ eni_20.png?​direct&​300 |Tighten the horizontal adjustment screws after alignment.}}]
  
 **Note: there should be no point throughout the 360º rotation of the fine focus knob where an increase in drag is felt.  If drag is felt, repeat the above steps. ​  In some cases sliding the flex coupler back and forth between the drive shaft and the Drive Gear shaft will help determine if the two shafts are properly aligned – be sure to retighten the coupler set screws so there is no slippage!** **Note: there should be no point throughout the 360º rotation of the fine focus knob where an increase in drag is felt.  If drag is felt, repeat the above steps. ​  In some cases sliding the flex coupler back and forth between the drive shaft and the Drive Gear shaft will help determine if the two shafts are properly aligned – be sure to retighten the coupler set screws so there is no slippage!**
Line 109: Line 109:
 ===== Part 5:  Installing the Motor Drive Cover Plate and Fine Focus Knob ===== ===== Part 5:  Installing the Motor Drive Cover Plate and Fine Focus Knob =====
  
-[{{ :​documentation:​eni_21.png?​direct&​300 |Install cover with screws as shown}}]+[{{ eni_21.png?​direct&​300 |Install cover with screws as shown}}]
  
 1. Locate the motor drive cover. ​ Position it over the motor drive assembly and secure in place as shown, using the two small cover screws with the 2mm hex wrench provided. 1. Locate the motor drive cover. ​ Position it over the motor drive assembly and secure in place as shown, using the two small cover screws with the 2mm hex wrench provided.
Line 121: Line 121:
 5. Press the knob on till it magnetically locks onto the Drive Plate. 5. Press the knob on till it magnetically locks onto the Drive Plate.
  
-[{{ :​documentation:​eni_22.png?​direct&​300 |Slide on the drive plate over the shaft of the Z-Drive.}}]+[{{ eni_22.png?​direct&​300 |Slide on the drive plate over the shaft of the Z-Drive.}}]
               ​               ​
-{{ :​documentation:​eni_23.png?​direct&​300 |}}+{{ eni_23.png?​direct&​300 |}}
  
-[{{ :​documentation:​eni_24.png?​direct&​300 |Ensure the Drive plate is flush with the end of the drive shaft. ​ Secure it with the two set screws using a  1.5 mm hex wrench.}}]+[{{ eni_24.png?​direct&​300 |Ensure the Drive plate is flush with the end of the drive shaft. ​ Secure it with the two set screws using a  1.5 mm hex wrench.}}]
  
 Align the posts on the original microscope knob the slots on the drive plate and press on.   The knob should magnetically attach to the plate, completing the installation procedure. Align the posts on the original microscope knob the slots on the drive plate and press on.   The knob should magnetically attach to the plate, completing the installation procedure.
  
-{{ :​documentation:​eni_25.png?​direct&​300 |}}+{{ eni_25.png?​direct&​300 |}}
  
-{{ :​documentation:​eni_26.png?​direct&​300 |}}+{{ eni_26.png?​direct&​300 |}}
  
-{{ :​documentation:​eni_27.png?​direct&​300 |}}+{{ eni_27.png?​direct&​300 |}}
  
  