+ | ====== Olympus IX70 Linear Encoder Installation ====== | ||
+ | The procedure outlines the steps necessary to install and align the ASI Heidenhain linear encoder onto the Olympus IX70 series microscopes. | ||
+ | |||
+ | - Installing the encoder clamp | ||
+ | - Installing and aligning the encoder | ||
+ | |||
+ | The procedure requires the following Allen wrenches that are supplied with the unit: | ||
+ | |||
+ | * 2.5mm | ||
+ | * 3/32” | ||
+ | |||
+ | |||
+ | ===== Step # 1 Installing the encoder clamp ===== | ||
+ | | ||
+ | |||
+ | [{{ :: | ||
+ | | ||
+ | [{{ :: | ||
+ | |||
+ | To install the linear encoder the original stage must first be removed; please refer to your microscope’s manual if needed. | ||
+ | |||
+ | |||
+ | ===== Step # 2 Installing & Aligning the Linear Encoder ===== | ||
+ | |||
+ | |||
+ | Please note: If you purchased the MS-2000-XYZ-LE unit with an automated XY stage the plunger will engage with the rear lead screw cover on the XY stage. | ||
+ | |||
+ | [{{ :: | ||
+ | |||
+ | [{{ :: | ||
+ | |||
+ | [{{ :: | ||
+ | |||
+ | |||
+ | Locate the Heidenhain encoder and the 3/32 inch allen wrench. Use the Allen wrench to insure that the screw on the side of the encoder clamp is loose. | ||
+ | |||
+ | Please note that the encoder has a total travel of 12 mm and that it should be positioned to allow the most convenient travel distance for the stage. In most instances the downward movement of the objective of has the longest travel distance. | ||
+ | |||
+ | <WRAP center round important 60%> | ||
+ | **WARNING!** Please do not move the stage outside of the linear encoder’s range without first disengaging the drive, selecting the rotary encoder, or removing power from the controller. | ||
+ | </ | ||
+ | |||
+ | |||
+ | This completes the installation and alignment of the ASI linear encoder onto the Olympus IX70 microscope. | ||
+ | |||
+ | {{tag> |