The ASI Stable Stage plugin can measure and correct thermal drift.

Requirements: Micro-Manager 2.0

Optional: CRISP or Autofocus3

Main Tab:

1) Run Forever: Check this box if you would like to run the test indefinitely in Drift Measurement or Stage Stabilization mode.

2) Test Length: If you're not running the test indefinitely, then you can specify the number of minutes you would like the test to run.

3) Drift Error: The amount of drift in microns required for Stage Stabilization to attempt to correct the XY stage position.

4) Max Move: The maximum amount that Stage Stabilization will move the stage in microns.

5) Phase Correlation: Which phase correlation algorithm to use for Stage Stabilization.

6) Autofocus Device: Which autofocus device to use for Stage Stabilization. Select None if no device available.

7) XY Moves: Invert the computed shift values in either the X or Y axes.

8) XY Gain: A multiplier to adjust the computed shift value.