 ===== Configuring PMT-200 to operate with 15V or 5V PMTs ===== ===== Configuring PMT-200 to operate with 15V or 5V PMTs =====
  
-Some PMT modules like H10722 operate with +/- 5Volts, and others like H9306 operate with +/- 15Volts. ​TGPMT card is compatible with both PMTs, however it needs to be configured with Jumper to provide the necessary voltage to the PMTs. +Some PMT modules like H10722 operate with +/- 5Volts, and others like H9306 operate with +/- 15Volts. ​PMT-200 ​is compatible with both PMTs, however it needs to be configured with Jumper to provide the necessary voltage to the PMTs. 
  
 Note the configuration is card wide and applies to both channels. Two PMTs with different input voltages cannot be used with the PMT-200 card at the same time.  Note the configuration is card wide and applies to both channels. Two PMTs with different input voltages cannot be used with the PMT-200 card at the same time. 
Line 74: Line 74:
 [{{ pmt_rj45.jpg?​direct&​300 |[[https://​en.wikipedia.org/​wiki/​TIA/​EIA-568|RJ45 Ethernet cable in EIA-568B confg]]}}] [{{ pmt_rj45.jpg?​direct&​300 |[[https://​en.wikipedia.org/​wiki/​TIA/​EIA-568|RJ45 Ethernet cable in EIA-568B confg]]}}]
  
-TGPMT card connects to the PMT tubes thru a [[https://​en.wikipedia.org/​wiki/​TIA/​EIA-568|RJ45 Ethernet cable in EIA-568B configuration]]. This is common cable used in IT/​networking. If your cable is malfunctioning or misplaced, in a pinch a common ethernet cable could be used instead. If in doubt contact ASI for a replacement. ​+PMT-200 ​connects to the PMT tubes thru a [[https://​en.wikipedia.org/​wiki/​TIA/​EIA-568|RJ45 Ethernet cable in EIA-568B configuration]]. This is common cable used in IT/​networking. If your cable is malfunctioning or misplaced, in a pinch a common ethernet cable could be used instead. If in doubt contact ASI for a replacement. ​
  
 [{{ pmt_connected.jpg?​direct&​300 |TGPMT card shown connected to two PMTs}}] [{{ pmt_connected.jpg?​direct&​300 |TGPMT card shown connected to two PMTs}}]
  
 <WRAP center round important 60%> <WRAP center round important 60%>
-The connector on the PMT and TGPMT card are **NOT** ethernet connector. They are wired in a different manner. Connecting anything other than a PMT to the TGPMT card, or vice-versa will cause damage and failure. ​+The connector on the PMT and PMT-200 ​are **NOT** ethernet connector. They are wired in a different manner. Connecting anything other than a PMT to the PMT-200, or vice-versa will cause damage and failure. ​
 </​WRAP>​ </​WRAP>​
  