tuning_40_tpi_stage
Differences
This shows you the differences between two versions of the page.
Previous revision | |||
— | tuning_40_tpi_stage [2021/09/23 17:15] (current) – external edit 127.0.0.1 | ||
---|---|---|---|
Line 1: | Line 1: | ||
+ | ====== Tuning Your 40 TPI Lead screw Stage for Small Moves ====== | ||
+ | |||
+ | Dear User, your stage has been optimized for small move sizes (100nm-10nm). Various physical and firmware settings were modified to accomplish this. To achieve 10nm moves we found the following settings helped. | ||
+ | |||
+ | <asi> $ Kp x=250 y=250 | ||
+ | $ Ki x=5 y=5 | ||
+ | $ Ma x=2 y=2</ | ||
+ | |||
+ | While these settings will help for submicron moves, they may cause instability and jitter. If you notice this feel free to revert one or more of these settings back to their defaults. | ||
+ | |||
+ | First, <wrap cmd> Ma x=0 y=0 </ | ||
+ | |||
+ | And if the instability and jitter persists | ||
+ | |||
+ | <asi> $ Kp x=75 y=75 | ||
+ | $ Ki x=1 y=1</ | ||
+ | |||
+ | |||
+ | If you have further questions please contact us at support[at]asiimging[dot]com or thru 1-541-461-8181 | ||
+ | |||
+ | |||
+ | |||
+ | {{tag> |
Address: 29391 W Enid Rd. Eugene, OR 97402, USA | Phone: +1 (541) 461-8181