+ | ====== Zeiss AXIOVERT-35 | ||
+ | (left side drive) | ||
+ | |||
+ | The instruction steps below describe how to install the ASI motor drive on to the Zeiss Axiovert-35 Microscope. | ||
+ | |||
+ | == The procedure has three parts: == | ||
+ | |||
+ | 1. Removing the plastic bottom cover from the underside of the microscope and installing in its place the motor drive base plate. | ||
+ | |||
+ | 2. Installing the motor drive assembly on to the base plate and microscope. | ||
+ | |||
+ | 3. Adjusting the motor drive. | ||
+ | |||
+ | The following tools will be required to perform this installation: | ||
+ | |||
+ | * 14mm deep socket wrench | ||
+ | * 7/64 inch hex wrench | ||
+ | * 1/16” hex wrench | ||
+ | * 3mm hex wrench | ||
+ | * .050 inch hex wrench | ||
+ | * medium slotted screw driver | ||
+ | * small slotted screw driver | ||
+ | * medium phillips screw driver | ||
+ | |||
+ | |||
+ | ===== Part 1 - Installing the base plate ===== | ||
+ | |||
+ | |||
+ | 1. This step requires that the microscope be inverted. | ||
+ | |||
+ | 2. Unscrew the six screws securing the plastic bottom cover. | ||
+ | |||
+ | 3. Locate the ASI base plate. | ||
+ | |||
+ | 4. Re-invert the microscope by carefully picking it up and turning it over. | ||
+ | |||
+ | ===== Part 2 - Installing the drive assembly ===== | ||
+ | |||
+ | |||
+ | 1. Remove the left fine focus knob from the microscope by loosening the set screw in the side of the knob using the .050 inch hex wrench, then slide the knob off the shaft. | ||
+ | |||
+ | 2. Remove the left coarse focus knob by first removing the nut using the 14mm deep socket provided, then unscrew the coarse focus knob. The nut and knob turn counter-clockwise to unscrew. | ||
+ | |||
+ | 3. Locate the anti-backlash gear included with the ASI motor drive components. | ||
+ | |||
+ | 4. Loosen the screw in the rectangular block attached to the base plate so that the block can slide easily in the grove. | ||
+ | |||
+ | 5. Rotate the anti-backlash gear so that the two blackened teeth are visible. | ||
+ | |||
+ | 6. Now, position the drive assembly so that the drive shaft slips over the microscope fine focus shaft. | ||
+ | |||
+ | 7. Next, rotate the drive assembly slightly until the edge groves at the bottom towards the microscope are aligned with the mating surfaces on rectangular block. | ||
+ | |||
+ | 8. Insert and thread in the vertical adjustment screw, but do not tighten it yet. | ||
+ | |||
+ | |||
+ | ===== Part 3 - Aligning the drive assembly ===== | ||
+ | |||
+ | |||
+ | 1. Align the drive assembly to the microscope by carefully sliding it up and down, forwards and backwards slightly, while rotating the right fine focus knob. Alignment will be optimum when minimal friction is felt as the right fine focus knob is turned. | ||
+ | |||
+ | 2. Press in on the drive shaft to make sure it is fully over the microscope fine focus shaft. | ||
+ | |||
+ | 3. Remove the three painted screws located in the three edges of the base plate. | ||
+ | |||
+ | 4. Slide the microscope fine focus knob over the shaft extension. | ||
+ | |||
+ | The eye pieces, illuminators, | ||
+ | |||
+ | {{tag> | ||