 Remove the left fine focus knob from the microscope as follows: Remove the left fine focus knob from the microscope as follows:
  
-[{{ :​documentation:​zeiss_axioimag_zdrive_14_.jpeg?​direct&​300 |Remove rubber boot}}]+[{{ zeiss_axioimag_zdrive_14_.jpeg?​direct&​300 |Remove rubber boot}}]
  
-[{{ :​documentation:​zeiss_axioimag_zdrive_12_.jpeg?​direct&​300 |    Use 1.5 mm Allen wench to loosen set screw.}}]+[{{ zeiss_axioimag_zdrive_12_.jpeg?​direct&​300 |    Use 1.5 mm Allen wench to loosen set screw.}}]
  
-[{{ :​documentation:​zeiss_axioimag_zdrive_13_.jpeg?​direct&​300 |Pull knob straight off to expose shaft}}]+[{{ zeiss_axioimag_zdrive_13_.jpeg?​direct&​300 |Pull knob straight off to expose shaft}}]
  
 Remove rubber boot Use 1.5 mm Allen wench to loosen set screw. Pull knob straight off to expose shaft. Remove rubber boot Use 1.5 mm Allen wench to loosen set screw. Pull knob straight off to expose shaft.
Line 38: Line 38:
  
  
-[{{ :​documentation:​zeiss_axioimag_zdrive_2_.jpeg?​direct&​300 |Figure 4a}}]+[{{ zeiss_axioimag_zdrive_2_.jpeg?​direct&​300 |Figure 4a}}]
  
-[{{ :​documentation:​zeiss_axioimag_zdrive_15_.jpeg?​direct&​300 |Figure 4b: Uninstalled baseplate & Drive }}]+[{{ zeiss_axioimag_zdrive_15_.jpeg?​direct&​300 |Figure 4b: Uninstalled baseplate & Drive }}]
  
-{{ :​documentation:​zeiss_axioimag_zdrive_16_.jpeg?​direct&​300 |Figure 5: Installed baseplate }}+{{ zeiss_axioimag_zdrive_16_.jpeg?​direct&​300 |Figure 5: Installed baseplate }}
  
 Locate the baseplate / Drive assembly and remove the drive from the base plate if it is attached by removing the vertical adjustment screw shown in figure 4a, also loosen, but do not remove the horizontal adjustment screw shown in figure 4b. Lift the front of the microscope up and slide the baseplate under the microscope so that the area where the shaft was at is approximately centered under the left focus knob as shown in figure # 5. Locate the baseplate / Drive assembly and remove the drive from the base plate if it is attached by removing the vertical adjustment screw shown in figure 4a, also loosen, but do not remove the horizontal adjustment screw shown in figure 4b. Lift the front of the microscope up and slide the baseplate under the microscope so that the area where the shaft was at is approximately centered under the left focus knob as shown in figure # 5.
Line 48: Line 48:
 ==== Part-3 Installing & Aligning the Motor Drive ==== ==== Part-3 Installing & Aligning the Motor Drive ====
  
-{{ :​documentation:​zeiss_axioimag_zdrive_17_.jpeg?​direct&​300 |Figure 7a}}+{{ zeiss_axioimag_zdrive_17_.jpeg?​direct&​300 |Figure 7a}}
  
-{{ :​documentation:​zeiss_axioimag_zdrive_18_.jpeg?​direct&​300 |Figure 7b: Slide ASI dive shaft onto microscope’s fine focus shaft}}+{{ zeiss_axioimag_zdrive_18_.jpeg?​direct&​300 |Figure 7b: Slide ASI dive shaft onto microscope’s fine focus shaft}}
  
 Locate the ASI motor drive. Install the drive by using the 7/64” Allen wrench to loose the clamp at the end of the ASI dive shaft. This clamp is located near the black encoder case. After loosening the clamp slide the ASI drive shaft over the microscope’s fine focus shaft, as shown in figure # 7, and then slide the tab on the bottom of the adjustment bar into slot on the base plate (Please Note the baseplate may need to be moved to properly align the adjustment bar.) Use the 3mm horizontal adjustment screw removed in part 2 to secure the drive to the base plate, as shown in figure #8 below. Leave the screw loose enough so that the adjustment bar can slide easily in the groove in the baseplate. Use the 7/64 inch hex wrench to tighten the drive shaft clamp, as shown in figure #9 below. Locate the ASI motor drive. Install the drive by using the 7/64” Allen wrench to loose the clamp at the end of the ASI dive shaft. This clamp is located near the black encoder case. After loosening the clamp slide the ASI drive shaft over the microscope’s fine focus shaft, as shown in figure # 7, and then slide the tab on the bottom of the adjustment bar into slot on the base plate (Please Note the baseplate may need to be moved to properly align the adjustment bar.) Use the 3mm horizontal adjustment screw removed in part 2 to secure the drive to the base plate, as shown in figure #8 below. Leave the screw loose enough so that the adjustment bar can slide easily in the groove in the baseplate. Use the 7/64 inch hex wrench to tighten the drive shaft clamp, as shown in figure #9 below.
  
-{{ :​documentation:​zeiss_axioimag_zdrive_2_.jpeg?​direct&​300 | Figure 8: Install but do no tighten horizontal adjustment screw}}+{{ zeiss_axioimag_zdrive_2_.jpeg?​direct&​300 | Figure 8: Install but do no tighten horizontal adjustment screw}}
  
-{{ :​documentation:​zeiss_axioimag_zdrive_1_.jpeg?​direct&​300 | Figure 9: Tightening the drive shaft clamp}}+{{ zeiss_axioimag_zdrive_1_.jpeg?​direct&​300 | Figure 9: Tightening the drive shaft clamp}}
  
 Slide the motor drive up and down; forward and backward //​slightly//​ while turning the right fine focus knob until it is in the position where minimum drag is felt. Secure the motor drive into position by tightening the horizontal and vertical adjustment screws loosened in step #2, figure 4 a & b. Use the 0.05” Allen wrench to tighten the baseplate clamp shown in figure # 10 below. NOTE: The baseplate clamp should be tightened just enough to secure it in place. Slide the motor drive up and down; forward and backward //​slightly//​ while turning the right fine focus knob until it is in the position where minimum drag is felt. Secure the motor drive into position by tightening the horizontal and vertical adjustment screws loosened in step #2, figure 4 a & b. Use the 0.05” Allen wrench to tighten the baseplate clamp shown in figure # 10 below. NOTE: The baseplate clamp should be tightened just enough to secure it in place.
Line 62: Line 62:
 Recheck the alignment by noting the drag on the right fine focus knob. No noticeable drag should be felt Repeat the steps above if necessary. Recheck the alignment by noting the drag on the right fine focus knob. No noticeable drag should be felt Repeat the steps above if necessary.
  
-{{ :​documentation:​zeiss_axioimag_zdrive_3_.jpeg?​direct&​300 |Figure 10 Tighten baseplate clamp after drive is installed & roughly aligned}}+{{ zeiss_axioimag_zdrive_3_.jpeg?​direct&​300 |Figure 10 Tighten baseplate clamp after drive is installed & roughly aligned}}
  
 ==== Part 4-Installing the motor drive cover plate & fine focus knob ==== ==== Part 4-Installing the motor drive cover plate & fine focus knob ====
Line 68: Line 68:
 Locate the motor drive cover. Loosen the 4/40 button head screws from the drive with the 1/16 “ Allen wrench. Position the motor drive covert over the motor drive assembly and secure in place using the 4/40 button head screws. Locate the motor drive cover. Loosen the 4/40 button head screws from the drive with the 1/16 “ Allen wrench. Position the motor drive covert over the motor drive assembly and secure in place using the 4/40 button head screws.
  
-[{{ :​documentation:​zeiss_axioimag_zdrive_5_.jpeg?​direct&​300 |Figure 11a}}]+[{{ zeiss_axioimag_zdrive_5_.jpeg?​direct&​300 |Figure 11a}}]
  
-[{{ :​documentation:​zeiss_axioimag_zdrive_4_.jpeg?​direct&​300 |Figure 11b}}]+[{{ zeiss_axioimag_zdrive_4_.jpeg?​direct&​300 |Figure 11b}}]
  
-[{{ :​documentation:​zeiss_axioimag_zdrive_7_.jpeg?​direct&​300 |Figure 11c}}]+[{{ zeiss_axioimag_zdrive_7_.jpeg?​direct&​300 |Figure 11c}}]
  
-[{{ :​documentation:​zeiss_axioimag_zdrive_6_.jpeg?​direct&​300 |Figure 11d: Installing the cover }}]+[{{ zeiss_axioimag_zdrive_6_.jpeg?​direct&​300 |Figure 11d: Installing the cover }}]
  
 Slide the microscope fine focus knob over the shaft extension and secure in place using the 1.5mm Allen wrench. Please note that the setscrew in the knob should mate with the flat on the shaft. Slide the microscope fine focus knob over the shaft extension and secure in place using the 1.5mm Allen wrench. Please note that the setscrew in the knob should mate with the flat on the shaft.
Line 80: Line 80:
 ===== Installation of the ASI Heidenhain linear encoder onto the Zeiss AxioImager Microscope ===== ===== Installation of the ASI Heidenhain linear encoder onto the Zeiss AxioImager Microscope =====
  
-{{ :​documentation:​zeiss_axioimag_zdrive_8_.jpeg?​direct&​300 |}}+{{ zeiss_axioimag_zdrive_8_.jpeg?​direct&​300 |}}
  
 The procedure outlines the steps necessary to install and align the ASI Heidenhain linear encoder onto the Zeiss AxioImager microscope. The linear encoder mounts to the rear of the microscope via an encoder clamp. The linear encoder has a plunger that depresses into the encoder as the focusing position is moved. The plunger tip mates with a plunger stop that is attached to the microscopes stage carrier. The linear encoder installation has two parts: The procedure outlines the steps necessary to install and align the ASI Heidenhain linear encoder onto the Zeiss AxioImager microscope. The linear encoder mounts to the rear of the microscope via an encoder clamp. The linear encoder has a plunger that depresses into the encoder as the focusing position is moved. The plunger tip mates with a plunger stop that is attached to the microscopes stage carrier. The linear encoder installation has two parts:
Line 91: Line 91:
 ==== Step # 1 Installing the plunger stop ==== ==== Step # 1 Installing the plunger stop ====
  
-{{ :​documentation:​zeiss_axioimag_zdrive_9_.jpeg?​direct&​300 |Figure 1a.Plunger stop mounts to stage carrier with the two screws as shown. }}+{{ zeiss_axioimag_zdrive_9_.jpeg?​direct&​300 |Figure 1a.Plunger stop mounts to stage carrier with the two screws as shown. }}
  
 ==== Step # 2 Installing the encoder clamp & Aligning the encoder ==== ==== Step # 2 Installing the encoder clamp & Aligning the encoder ====
  
-{{ :​documentation:​zeiss_axioimag_zdrive_10_.jpeg?​direct&​300 |Figure 2a Plunger stop mounts to stage carrier with the two screws as shown.}}+{{ zeiss_axioimag_zdrive_10_.jpeg?​direct&​300 |Figure 2a Plunger stop mounts to stage carrier with the two screws as shown.}}
  
-{{ :​documentation:​zeiss_axioimag_zdrive_11_.jpeg?​direct&​300 |Figure 3a Aligning & securing linear encoder }}+{{ zeiss_axioimag_zdrive_11_.jpeg?​direct&​300 |Figure 3a Aligning & securing linear encoder }}
  
 Bring the stage up to the correct focal position for the common objectives used. Position the encoder & encoder clamp so that the ball on the end of the encoder’s plunger mates with the triangular carbide plunger stop as shown in figure 3a. Insert the encoder into the encoder holder & slide the encoder down until there is about 1 to 3 millimeters worth of upward travel left on the encoder’s plunger. Hold the encoder in place & use the 3/32 inch Allen wrench to tighten the screw to secure the encoder in place. Bring the stage up to the correct focal position for the common objectives used. Position the encoder & encoder clamp so that the ball on the end of the encoder’s plunger mates with the triangular carbide plunger stop as shown in figure 3a. Insert the encoder into the encoder holder & slide the encoder down until there is about 1 to 3 millimeters worth of upward travel left on the encoder’s plunger. Hold the encoder in place & use the 3/32 inch Allen wrench to tighten the screw to secure the encoder in place.